The fluxmeter FG16 is a modular device to measure magnetic values and can be equipped with up to 4 channels.
The completely redesigned analog-/digital-hybrid design of the fluxmeter plug-in allows measurements of both high-dynamic as well as very fast events and is virtually drift-free.
The FG16 measures the following magnetic values:
- Flux Φ (in Wb, Vs)
- Flux density B (in T)
- Field strength H (in A/m)
- Polarization J (in T)
- Dipole moment m (in Am²)
Custom-designed probes complete the measuring application.
The FG16 is easy and intuitive to use with a touch display and separate buttons.
Additionally, the FG16 provides a variety of modern interfaces such as Ethernet, USB, CAN bus, RS232, analog and 24 V digital signals for control and output of measurement results.
The device is suited for demanding measuring applications in the field of research and development as well as for automatic process control, quality assurance and incoming goods inspection.
Typical application examples:
- Flux or flux density measurement of a work piece during magnetization
- Stray field measurement
- Test of the dipole moment and the polarization of a magnet
Probes with integrated parameter memory simplify correct measurements.
Comprehensive accessories, such as probes, reference magnets, also in special design, are available.
Special Features
- Modular design (up to 4 channels)
- Color touch display
- Easy to use
- Standard interfaces implemented
- Very low drift
- Probe connector with low thermo-electric potential
- Internal window comparator
- Automatic probe detection
- Self-test on power-up
- Integrated calibrator
- Digital pos./neg. peak memory
- Analog peak-hold function
- Firmware updates via PC
- 2nd input jack on rear side
- Frequency range 0 ... 20 kHz
- Sturdy case with option for 19 inch rack mount
- Measuring unit freely selectable
Specifications
Display | Color touch LCD, 95 mm × 52 mm, 480 × 270 pixels |
Display rate | 10 times per second |
Input connection | Jacks with low thermoelectric potential at front and rear side |
Unit systems | SI, CGS, support of metric and imperial units |
Interfaces | Analog ±10 V, 200 Ω RS232 DE-9 plug, adjustable parameters CAN bus DE-9 plug, adjustable parameters LAN RJ45, 10/100 Mbit 24 V IO DB-25 socket, 8 inputs, 8 outputs, optically decoupled USB 2.0, front side type A, rear side type B |
Dimensions | 125 mm × 260 mm × 260 mm (H × W× D) |
Weight | 2.85 kg (6.3 lbs.) incl. 4 plug-ins, without packaging and accessories |
Power supply | 100 – 240 V, 47 – 63 Hz, max. 25 VA, interference suppression for 50/60 Hz line frequency |
Temperature range | Operating: 0 °C to +55 °C < 80% relative humidity at +40 °C non-condensing Non-operating: −30 °C to +70 °C |
Fluxmeter Plug-in
Measuring method | Low-drift integrator with dynamic drift adjustment | |
Function/unit | Magnetic flux Φ | Wb, V·s, Mx |
Magnetic flux density B | T, G | |
Magnetic field strength H | A/m, Oe | |
Magnetic polarization J | T, G | |
Magnetic dipole moment m | A·m², V·s·m, erg/G | |
Magnetic potential difference Vm | A, Gb | |
Range (Resolution) |
Flux ± 1 mVs (0.1 μVs) |
The range and the resolution of the derived values depend on the used probe. |
Range selection | automatic, manual | |
Frequency range | 0 … 20 kHz | |
Accuracy | 0.5 % after self calibration | |
Calibration | Internal voltage and time reference | |
Drift | < ± 2 μVs/minute | |
Input resistance | 100 kΩ ± 0.1 % |